An analytical and Monte Carlo investigation of the sufficiency of the present shielding of PET/CT imaging system at Tehran's Shariati hospital
محل انتشار: فصلنامه فیزیک و مهندسی پرتو، دوره: 1، شماره: 3
سال انتشار: 1399
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 286
فایل این مقاله در 5 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_RPE-1-3_005
تاریخ نمایه سازی: 22 دی 1399
چکیده مقاله:
By the rapid development of imaging systems such as PET/CT for diagnosis of cancer, the protection of staff and public has become a main health concern. Due to serious and irreversible harms of ionization radiations, protection of all those who are exposed is the main concern of health issues. The main basis of the calculation of the shielding design in the medical imaging systems is that the absorbed dose should not exceed the allowed limit. In this study, the current shielding status of the PET/CT installations in Tehran's Shariati hospital was investigated using the MCNPX Monte Carlo code to ensure that the dose limits for both the controlled and uncontrolled area are not violated. The proposed simulation method was benchmarked with a validated analytical method. Shariati hospital provides services to four patients every day, leading to a dose rate in the range of 2.6 × 10-6 to 9.35 × 10-3 mSv/week. The minimum dose rate in this range represents the value behind the door of the waiting room (public uncontrolled area), while the maximum in this range corresponds to the value behind the glass of the scanner room (operator controlled area). The simulation results for 8 patients/day in this center showed that the dose rate behind the wall of the injection room will increase from 4.88 ×10-6 mSv/week to 2.81 × 10-2 mSv/week, which is well below the recommended levels. This indicates that the present shielding is adequate for up to four more patients per day.
کلیدواژه ها:
نویسندگان
Fereshteh Gholami
Faculty of Nuclear Engineering and Physics, Amirkabir University of Technology, Tehran, Iran.
Ehsan Alibeigi
Faculty of Nuclear Engineering and Physics, Amirkabir University of Technology, Tehran, Iran.
Mojtaba Shamsaei-Zafarghandi
Faculty of Nuclear Engineering and Physics, Amirkabir University of Technology, Tehran, Iran.
Ehsan Nazemi
Nuclear Science and Technology Research Institute, Tehran, Iran
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :