Clinical Validation of the Saadat Non-Invasive Blood Pressure Module According to the British Standard EN ISO ۸۱۰۶۰-۲ Protocol
عنوان مقاله: Clinical Validation of the Saadat Non-Invasive Blood Pressure Module According to the British Standard EN ISO ۸۱۰۶۰-۲ Protocol
شناسه ملی مقاله: JR_SPRE-5-1_003
منتشر شده در در سال 1400
شناسه ملی مقاله: JR_SPRE-5-1_003
منتشر شده در در سال 1400
مشخصات نویسندگان مقاله:
Mohammadamin Haghayegh - Najafabad Branch, Islamic Azad University
Neda Behzadfar - Najafabad Branch, Islamic Azad University
خلاصه مقاله:
Mohammadamin Haghayegh - Najafabad Branch, Islamic Azad University
Neda Behzadfar - Najafabad Branch, Islamic Azad University
This study was conducted according to the British Standard EN ISO ۸۱۰۶۰-۲ guidelines. The aim was to validate the Saadat non-invasive blood pressure (NIBP) module accuracy and reliability against manual auscultatory general practitioners (GP), readings. A total of ۲۹۸ measurements and comparison procedures were performed on ۹۵ adults without heart disease under the supervision of two GPs. In order to represent the relationships between the test device and the reference method, the Bland-Altman graphical plotting method was used. The mean differences and standard deviations (mean ± SD differences) between the readings of the Saadat NIBP module and determination of GPs as the reference method for systolic and diastolic were exceptionally close with: -۲.۲۲ ± ۶.۵۱ mmHg and -۳.۳۱±۶.۲۷ mmHg respectively. The Saadat NIBP module fulfilled the BS EN ISO ۸۱۰۶۰-۲ requirements, which states that the mean ± SD of lower than ۵±۸ mmHg can be recommended for clinical use.
کلمات کلیدی: Saadat NIBP, Validation, BS EN ISO ۸۱۰۶۰-۲, manual auscultatory