Optimization of Head CT Protocol to Reduce the Absorbed Dose in Eye Lenses and Thyroid: A Phantom Study

عنوان مقاله: Optimization of Head CT Protocol to Reduce the Absorbed Dose in Eye Lenses and Thyroid: A Phantom Study
شناسه ملی مقاله: JR_IJMP-16-1_010
منتشر شده در در سال 1398
مشخصات نویسندگان مقاله:

Parva Kalhor - Tehran University of medical Sciences school of Allied medical Sciences, Tehran province, Tehran, Keshavarz Blvd., Ghods Street, District ۶, Farre danesh
Vahid Changizi - Tehran University of Medical Sciences school of Allied Medical Sciences, Iran, Tehran Province, Tehran, Keshavarz BLVD., Ghods Street, District ۶, Farre danesh
Seyed Abolfazl Hosseini - Sharif University of Technology Department of Energy Engineering, Iran, Tehran province, Tehran, Azadi Street
Elahe Jazayeri - Tehran University of Medical Sciences school of Allied Medical Sciences, Iran, Tehran Province, Tehran, Keshavarz BLVD., Ghods Street, District ۶, Farre danesh

خلاصه مقاله:
Introduction: Utilization of computed tomography (CT) scans is increasing annually. This study aimed to reduce the absorbed dose of sensitive organs in the head (eye lenses and thyroid) and to assess changes in resultant images quality in head scans when the radiation dose is decreased. Material and Methods: An anthropomorphic phantom was examined with head protocols in both helical and sectional modes using two ۱۶-slice CT scanners. The entrance surface dose of eye lenses and thyroid was measured with standard protocols and after reducing the mAS and kilo-voltage using thermo-luminescence dosimeters (TLDs). Results: In sectional mode with standard protocol, the highest surface dose was ۲.۳ mSv۱ for eye lens and ۰.۰۲۱ mSv for thyroid in hospital A. Moreover, in helical mode with standard protocol, the highest surface dose was ۰.۹۶۴ and ۰.۰۲ mSv for eye lens and thyroid in hospital B, respectively. Reducing tube current and kilovoltage decreased the dose up to ۳۵% for eye lens and ۴۵% for thyroid in hospital A. By the mentioned reductions a dose decrease of up to ۴۰% was achieved for both eye lens and thyroid in hospital B. There were no considerable differences in image quality between scans with standard protocol and the protocols of reduced parameters. Conclusion: Head CT scans with standard factors conduce to images with the best quality. It may be possible to diminish the absorbed dose up to ۴۰% without losing information, especially in follow up head scans.

کلمات کلیدی:
Computed Tomography (CT), Eye Lens, Thyroid