Energy Window Optimization Using Triple Energy Window Scatter Correction Method for In-۱۱۱ SPECT Imaging: A SIMIND Monte Carlo Study
محل انتشار: مجله فیزیک پزشکی ایران، دوره: 18، شماره: 5
سال انتشار: 1400
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 174
فایل این مقاله در 6 صفحه با فرمت PDF قابل دریافت می باشد
- صدور گواهی نمایه سازی
- من نویسنده این مقاله هستم
استخراج به نرم افزارهای پژوهشی:
شناسه ملی سند علمی:
JR_IJMP-18-5_001
تاریخ نمایه سازی: 1 آبان 1400
چکیده مقاله:
Introduction: Detecting scattered photons in the photo peak window degrades the image contrast and quantitative accuracy of single-photon emission computed tomography (SPECT) imaging. This study aimed to determine optimal main- and sub-energy windows for Triple Energy Window (TEW) in In-۱۱۱. Material and Methods: We used the simulating medical imaging nuclear detectors (SIMIND) program to simulate the Siemens SYMBIA gamma camera equipped with a medium energy (ME) collimator. We also used the SIMIND Monte Carlo program to generate theIn-۱۱۱SPECT projection data of the Jaszczak phantom. The phantom consisting of six spheres with different diameters (۹.۵, ۱۲.۷, ۱۹.۱, ۱۵.۹, ۲۵.۴, and ۳۱.۸ mm) was used to evaluate the image contrast. Geometric, scatter, and penetration fraction, point spread functions, and contrast curves were drawn and compared. Results: The results showed that the ۱۷۱keVphotopeak compared to the ۲۴۵keVphotopeak yielded the best results with an ME collimator when the TEW scatter correction method was applied. The reason can be the large amount of scatter and penetration from the photo peak and the collimator for the ۲۴۵keVphotopeak window. Conclusion: With the TEW scatter correction method, it is better to use a ۱۷۱keVphotopeak window because of its better spatial resolution and image contrast.
کلیدواژه ها:
نویسندگان
Hicham Asmi
Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco
Farida Bentayeb
Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco
Youssef Bouzekraoui
Higher Institute of Health Sciences Settat, University HASSAN, Rabat, Morocco
Faustino Bonutti
Academic Hospital of Udine, Medical Physics Department, Italy
مراجع و منابع این مقاله:
لیست زیر مراجع و منابع استفاده شده در این مقاله را نمایش می دهد. این مراجع به صورت کاملا ماشینی و بر اساس هوش مصنوعی استخراج شده اند و لذا ممکن است دارای اشکالاتی باشند که به مرور زمان دقت استخراج این محتوا افزایش می یابد. مراجعی که مقالات مربوط به آنها در سیویلیکا نمایه شده و پیدا شده اند، به خود مقاله لینک شده اند :