Effects of Rotational Motion in Robotic Needle Insertion

سال انتشار: 1394
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 50

فایل این مقاله در 10 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

JR_JBPE-5-4_007

تاریخ نمایه سازی: 30 دی 1402

چکیده مقاله:

Background: Robotic needle insertion in biological tissues has been known as one the most applicable procedures in sampling, robotic injection and different medical therapies and operations.Objective: In this paper, we would like to investigate the effects of angular velocity in soft tissue insertion procedure by considering force-displacement diagram. Non-homogenous camel liver can be exploited as a tissue sample under standard compression test with Zwick/Roell device employing ۱-D axial load-cell.Methods: Effects of rotational motion were studied by running needle insertion experiments in ۵, ۵۰ and ۲۰۰ mm/min in two types of with or without rotational velocity of ۵۰, ۱۵۰ and ۳۰۰ rpm. On further steps with deeper penetrations, friction force of the insertion procedure in needle shaft was acquired by a definite thickness of the tissue.Results: Designed mechanism of fixture for providing different frequencies of rotational motion is available in this work. Results for comparison of different force graphs were also provided.Conclusion: Derived force-displacement graphs showed a significant difference between two procedures; however, tissue bleeding and disorganized micro-structure would be among unavoidable results.Â

نویسندگان

H Ramezanpour

BSc Student of Biomedical Engineering in Amirkabir University of Technology