Static Coil Design Considerations for the Magnetic Resonance Imaging

سال انتشار: 1397
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 294

فایل این مقاله در 7 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

JR_IJE-32-3_006

تاریخ نمایه سازی: 10 آذر 1398

چکیده مقاله:

One of the main challenges in developing magnetic resonance imaging (MRI) systems is to create a static coil that needs to generate magnetic field density along with the characteristics of optimal homogeneity and magnitude size. To do this, two N42 Block PMs are used and the iron core is designed and optimized in accordance with the dimensions of PM pieces using ANSYS Maxwell software. Then, all iron parts are lathed, the yoke pieces and pole spacers are welded. In addition, PM and pole pieces are installed. Finally, measurement is done by Lutron to evaluate the static coil performance.

نویسندگان

M. Shiravi

Department of Electrical & Computer Engineering, University of Kashan, Kashan, Iran

B. Ganji

Department of Electrical & Computer Engineering, University of Kashan, Kashan, Iran

A. Shiravi

Department of Radiology, Faculty of Medicine, Tehran University of Medical Sciences, Tehran, Iran