Modeling Signals Received from the Circular Concentrator Transducer in Non-Destructive Ultrasonic Test

سال انتشار: 1399
نوع سند: مقاله کنفرانسی
زبان: انگلیسی
مشاهده: 274

فایل این مقاله در 10 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

MMMC01_008

تاریخ نمایه سازی: 18 اردیبهشت 1400

چکیده مقاله:

To obtain correct results from ultrasound tests, the ultrasonic measurement system must be calibrated. To this end, base reflectors including side holes are used for calibration purposes. Besides, to enhance the efficiency of ultrasonic tests, usually, concentrating transducers are used. Therefore, the existence of a theoretical model for the reduction of the ultrasonic signals received from a base reflector through a concentrating transducer is considered both beneficial and significant. To model the ultrasonic inspection system, an appropriate model must be offered for any of the three parts of the system. They include ۱) transducer's ultrasonic wavefield; ۲) system function and ۳) waves' scattering field from the side-drilled holes. To model wave fields, a multi-Gaussian beam model is used to model system function from the signal received from the outer surface of the calibration clock. Besides, to model the scattering field, Kirchhoff approximation is used. Finally, the accuracy of the provided model has been validated using the experimental results.

نویسندگان

Amin Heydarian Hamzehkanloo

Department of Mechanical Engineering, Faculty of Montazeri, Khorasan Razavi Branch,Technical and Vocational University (TVU), Mashhad, Iran