Calculation of full-energy peak efficiency of NaI (Tl) detectors by new analytical approach for parallelepiped sources

سال انتشار: 1392
نوع سند: مقاله ژورنالی
زبان: انگلیسی
مشاهده: 315

فایل این مقاله در 11 صفحه با فرمت PDF قابل دریافت می باشد

استخراج به نرم افزارهای پژوهشی:

لینک ثابت به این مقاله:

شناسه ملی سند علمی:

JR_JTAP-7-1_002

تاریخ نمایه سازی: 27 مرداد 1397

چکیده مقاله:

A new analytical approach is presented for the calculation of full-energy peak (FEP) efficiency of NaI (Tl) detectors.The self-attenuation of the parallelepiped source matrix, the attenuation by the source container, and the detectorhousing materials were considered in the mathematical treatment. The efficiency values calculated using thepresently suggested analytical approach are compared with those measured values obtained by two different sizesof NaI (Tl) detectors. The calculated and the measured full-energy peak efficiency values were in a good agreement.

کلیدواژه ها:

نویسندگان

Ahmed M El-Khatib

Physics Department, Faculty of Science, Alexandria University, Alexandria ۲۱۵۱۱, Egypt

Mohamed S Badawi

Physics Department, Faculty of Science, Alexandria University, Alexandria ۲۱۵۱۱, Egypt

Mona M Gouda

Physics Department, Faculty of Science, Alexandria University, Alexandria ۲۱۵۱۱, Egypt